HIGH SPEED CUTTING INSTRUMENTS IN
PROSTHODONTICS
Introduction
In
order to perform the intricate and detailed procedures associated with
restorative dentistry, the dentist must have a complete knowledge of the
purpose and application of the many instruments required. During each day of
his clinical experience the dentist operates on vital tissues within the oral
cavity where a millimeter or a fraction there of, is a very significant
dimension. A skillful application of sharp hand and rotary instruments requires
ability and coordination gained only by extensive training.
Before
the advent of rotary instruments, removal of tooth tissue was accomplished with
sharp – edged chisels, hatchets, and hoes. These hand instruments possessed a
cutting capability, which was used for clearing away unsupported and undermined
enamel resulting from dental caries. Walls and floors of the cavity were formed
by a planning and lateral scraping action of these sharp edged instruments. At
best, such efforts were crude, time consuming and often difficult.
The
first, rotary instruments for cutting tooth tissue were modified hand
instruments. These, drill or bur heads could be twisted in the fingers to
produce a cutting or abrading action. In 1846 the finger ring was introduced
with a drill socket attached for adapting a series of long bundled burs or
drills. This was the primitive application of the rotary principle. The first
drill having flexible cable drive and the first angle hand piece were
introduced by Charles Merry between 1858 and 1862. In 1871, Morrison modified
and adapted the dental foot engine from the Singer Sewing machine. This was
followed by the introduction of the electric dental engine utilizing a cable
arm in 1883. In 1910 the endless cord on a jointed arm was made available. The
earlier dental hand pieces were capable of speeds from 4500 to 6000 rpm.
In
1940 the use of diamond abrasive paints became widespread. The diamond point is
compared of a number of small diamond particles bound on a rotary blank.
In
1945 Dr. G.V. black, published a report on the non mechanical preparation of
cavities and in doing so introduced the air abrasive technique. The impact of
Dr. Black’s revolutionary cutting technique on the dental profession was
considerable. This was the first significant break in the long established
traditional method of cavity preparation. The air abrasive principle utilized
particles of aluminium oxide propelled against the tooth surface by a carbon
dioxide stream under the pressure of 110 psi, and funneled through a tungsten
carbide nozzle with a lumen of 0.018 inch. The penetration of enamel and dentin
was rapid but some what difficult to control.
In
1949 Walsh and Symons published their initial findings relating to the removal
of tooth tissue with diamond points at rotational speeds upto 70,000 rpm. This
report indicated the use of lighter forces and a resulting increased cutting
efficiency at these higher speeds.
In
early 1950, the ball-bearing hand piece was introduced.
In
1963, following the work of Nelson the first fluid turbine type handpiece was
introduced. This instrument was capable of rotational speeds of approximately
50,000 rpm and was limited to diamond instruments operated at one speed only.
In 1954, air-driven hand pieces were developed. A continuous belt-driven
contra-angle which utilized a friction grip chuck and bur was introduced,
making possible cutting speeds of upto 150,000 rpm.
By 1957, many dentists were using rotational speeds upto 3,00,000
rpm. The introduction of air-bearing hand piece in the early 1950 made possible
greater rotational speeds of approximately 5,00,000 rpm.
In
1953, an ultrasonic method of tooth tissue
removal was also introduced, which used suitably shaped tips vibrating
at frequencies ranging from 2,50,000 to 3,00,000 cycles – per seconds.
This
brief historical back ground reveals that the profession has been searching for
a suitable method of tooth tissue removal. Only during last 30 years, this hunt
has slowed down still the profession is trying to refine the procedure and
instruments.
Review of literature
A
search through literature reveals various methods used in the past for removal
of tooth tissue. The continuous development of newer methods till 1960,
indicates that the earlier instruments
had some disadvantages. Inspite of the introduction of numerous tooth reduction
instruments, and procedures, the principles and the biologic objectives have
remained the same. These are as follows.
1.
The operator should remove the
least amount of tooth tissue consistent with necessary mechanical retention.
2.
This should be done with the least
barm to the periodontal tissues and the pulp.
3.
It should be done with the least
discomfort to the patient.
4.
No pathologic reactions should be
initiated in the pulp.
Advantages of high speeds
1.
Smaller stones can be used at the
increased speeds.
2.
Less fatigue results both for the
patient and operator.
3.
Due to high speed, very light
pressure is required.
4.
Less vibrations are felt by the
patient.
5.
The chairside time for a given
preparation is considerably reduced.
6.
Trauma to the pulp is reduced.
7.
The efficiency and life of the
cutting tools is increased.
8.
Because of small tools, control is
easy.
9.
Removal of old amalgam and gold
restorations is easy.
Disadvantages of high speeds
1.
The increased speed creates
increased temperatures in the tooth. Therefore some method of cooling the tooth
more efficiently is required not to injure the pulp. This necessitates
additional equipment.
2.
When a dentist changes from the
lower speeds, which utilize a pressure in pounds, to high speeds which need
only a pressure in ounces, he must develop a new technique and retrain himself
to a new tactile sense.
3.
To operate at high speeds good
visibility of the cutting instrument is necessary to avoid over cutting.
4.
Due to the ease with which tooth tissue
is removed, caution must be taken not to injure the proximal enamel of the
adjacent healthy tooth and the gingiva.
5.
High speeds result in greater wear
on the working parts of the handpiece, necessitating more frequent repairs and
replacements.
6.
Unless used properly, high speeds
have a tendency to create striations on a tooth surface.
7.
The ideal preparation for any type
of restoration cannot be accomplished by using high speed equipment alone. The
final exactness and finishing line can best be established by instruments
revolving at moderate speeds.
Types of high speed instruments
Hand
piece can be divided into four types depending upon their speeds as follows.
1.
Low speed – upto 10,000 rpm.
2.
Intermediate speed – 25,000 to
45,000 rpm.
3.
High speeds – 50,000 to 1,00,000
rpm.
4.
Ultra high speeds 1,00,000 rpm and
over.
Kilpatric has further classified
the ultra high speed handpiece into three classes.
Type I –
the gear driven centre-angle handpiece, upto1,25,000 rpm.
Type II –
the belt driven contra-angle handpiece upto 2,00,000 rpm.
Type III – turbine driven air
contra-angle handpiece upto 3,00,000 rpm and higher.
Heat generation:
Knowledge
of the physics tells us that, whenever there is friction between two surfaces,
heat is generated, which may bring about rise in temperature of either or both
the surfaces. The same applies in the tooth reduction procedures. Here the
rotating cutting tools come in contact with the tooth surface and the heat is
generated.
It
was not until 1930 that the workers began to investigate the heat rise in the
dental pulp.
There
are many factors that influence the rise in temperature which takes place in
cutting operations. The greater the speed of rotation of the cutting tool, the
faster the tool revolves, the higher the resultant temperature. It has been
found that the temperature rise develops within 10-12 seconds, after the
cutting operation is started. Size of the cutting instrument has an important
bearing on heat generation, since, its diameter affects the cutting speed at
its periphery. Larger the size of the cutting tool more the host generation.
A
third factor is the pressure applied by the dentist during cutting operation.
As the pressure increases, greater will be the rise in temperature.
Hudson
and associates in 1954 conducted a study on temperature developed in dental
cutting instruments from their study they have concluded that,
1.
The temperatures produced by dental
burs in cutting human dentin ranged from 125°F to 275°F. Since these
temperatures are above those, said to be tolerated by normal human dentin, it
would seem advisable to use some form of coolant.
2.
A significant decrease in time
required to accomplish a given operation is apparent, when high operating
speeds are used.
3.
The amounts of heat transferred to
the tooth from the bur decreases, at speeds above 12000 rpm, since cutting time
at these speeds is reduced and bur temperature remains.
Substantially constant and there is less heat trauma to the
vital structures.
Coolants:
From
the study of Hudson and Sweeney, it is evident that the temperatures reached
during tooth reduction procedures are above those said to be tolerated by
normal human dentins. This indicates that, some form of coolant must be used,
during the cutting operations, particularly when high speeds are used.
Every
means should be employed to keep the temperature down as much as possible
during cutting operations. Coolants must employed which, to be effective,
should be applied at the point of contact between the cutting instrument and
the tooth tissue. There are three types of coolants usually employed in dental
practice.
1. Water.
2. Spray of air and
water
3.
Air alone.
Peyton
has shown that at speeds ranging from 30000 rpm to 170000 rpm and with an
application of four ounces of pressure, a temperature rise within the tooth of
less than 15°C occurred when water or air-water sprays were employed. He also
found that even with a water coolant, excessive temperatures developed, when
large diameter instruments or excessive pressure were applied with increased
operating speeds. This indicates that the use of a coolant, does not eliminate the danger of
excessive temperature rise.
A
reduction in concentration of the amount of water used during cutting procedure
shows the significant temperature rise of the dental bur.
The
minimum volume of water to be applied was estimated at 1.5 ml per minute.
The
question whether water in spray form should be used at mouth or temperature
seems to have no significance as far as temperature rise in the tooth was
concerned. Tylman is of the opinion that if the water reservoir is kept at
100°F, it is most comfortable to the patient, less liable to be harmful to the
pulp and still reduces the heat of friction during cutting.
There
are certain other problems associated with the use of the highspeed cutting
tools. Most of the hand pieces are so designed that a spray or stream of water
is directed from the head of the handpiece directly onto the cutting operation.
Where the water strikes the tooth and the cutting tool directly, full benefit
is obtained from the coolant. Where however, the abrasive on the cutting tool,
is on the surface away from the stream of water, water does not flood the tooth
surface being cut, resulting in excessive temperature rise. The overcome this
difficulty perforated disks have been developed, which permit the water to go
through the openings and lubricate the disk and tooth on the cutting side. The
use of perforated disk results in less temperature rise. Consequently when
disks are non perforated, and when the stream of water cannot be directed to
the cutting contact areas, they should be used at speeds not exceeding 10,000
rpm.
Another
advantage of a water coolant lies in the fact that the tooth debris from the
cutting is removed rapidly, preventing the clogging of the cutting tools. This
results in greater cutting efficiency of the stone. Also, it prolongs the life
and effectiveness of the instrument. It is essential that the water be in
intimate contact with the revolving instrument and the tooth tissue. To do this more effectively, Nelson
recommended the addition of a wetting agent to the water spray.
Because
the high speed technique requires a larger quantity of water as a coolant,
there is the problem of removing this water from the mouth. To have the dentist
stop frequently to allow the patient to spit out the excess water is time
consuming. The customary saliva ejector has insufficient removal capacity.
To
solve this problem, Thompson has suggested a washed field technique.
This
technique adapts the suction or vacuum principle. It established and maintains
a powerful but gentle negative pressure of air in the mouth, close to the field
of operation.
Accompanying
the air stream, is a flow of isothermal water which is projected copiously onto
the operative field. This water is entrained into the vacuum air stream, which
draws it rapidly across the operative area. The irrigant pulls away with it
tooth cuttings and debris. These are taken into the vacuum air stream and
disposed off in a filter system. A clean, clearly visible operative field is
provided. This technique has the distinct advantages that it facilitates the
use of high speed instruments, maintains visibility during copious irrigation
of the operative field, reduces operating time, improves the patient’s well
being and introduces a new concept of cleanliness. Human tissues are maintained
in their natured wet safe pain, trauma and postoperative complication, which
may arise due to ingestion of tooth debris are reduced.
Desiccation
of hard and soft tissues is avoided. Heat is eliminated thus preserved the
tissues.
Vibration:
Cutting
a tooth may be very annoying and unpleasant to the patient but still not be
painful. In pain there is usually an involvement of the nerve endings, either
by trauma or extreme irritation, resulting in an acute, painful reaction. Most
patients associate the sensation of vibration, noise, pressure and the slight
increase in cutting temperature with the sensation of pain. Consequently, if
the factors of vibration, heat and pressure are reduced to a minimum, the
patient usually experience reduced or no pain.
One
mechanical factor that influences vibration is the dental handpiece, whether it
is friction-bearing, ball bearing, high speed belt driven or turbine ultra
speed. When the friction bearing, conventional type of handpiece is used at a
speed of 4500 rpm to 6000 rpm, it is connected by the conventional belt and
pulley system of the dental engine. In this case one may expect a high order of
vibration depending upon the condition of various mechanical parts, their
adjustment and speeds of their operation.
Pulleys
that are worn, a worn belt, or an improperly adjusted belt will cause
vibrations that are transmitted down to the cutting tool. Similarly hand piece
which do not hold the cutting tool properly, which have worm bearing or are out
of adjustment will also cause vibration.
The
investigations of Walsh and Symmoss showed that vibration, when applied to
tooth, produced the most unfavorable response when the frequency was between
100 cps and 200 cps. When the frequencies were above 1000 cps, they were
generally beyond the upper threshold of perception of the average patient. It
is the lower frequencies, in the range of 100 – 200 cps, that are usually
developed at the lower speeds, especially if the equipment is worm and
maladjusted.
Hudson
and Sweeney have reported the importance of having centricity in the cutting
tool. They found that eccentric burs when rotated at 6000 to 10000 rpm produced
a lower frequency in the range of 100-200 cps, whereas a true running bur at
10000 rpm produced vibrations in the frequency range above the upper threshold.
Tamner
pointed out that only a part of an eccentric cutting tool is used as it
rotates, thus causing unfavorable impacts and vibrations, which fall into the
most annoying frequency range. The disks and stones that are unmounted and are
screwed onto a mandrel very frequently are eccentric and therefore should not
be used in high speed cutting operations. The permanently mounted instruments
are indicated in preference to unmounted type.
Poorly
built burs with blades not evenly cut or chipped will likewise cause vibration.
In using carbide burs, it is very important the see that none have chipped
blades.
Correct
adjustment of the belt is important in the reduction on and elimination of
vibration. A belt that is too loose increases the vibration pattern transmitted
directly to the tooth.
In
the ultra highspeed hand pieces the metal chuck holding the cutting instrument
often is replaced by a rubber or plastic chuck. This lessens the vibration
transmitted to the cutting instrument and facilitates the more rapid cutting
action.
In
cutting with a water turbine handpiece at 45,000 rpm the intensity of
vibrations was well tolerates by the patient.
Morrison
and Grinnel made the following observations.
The
deleterious effects of vibration are two fold in origin.
1. Amplitude.
2.
Undesirable modulating frequencies.
If
we minimize or eliminate these factors, we can then reduce the undesirable
effects of vibration.
Amplitude: The wave of vibration consists of
frequency and amplitude.
At conventional speeds, amplitude
is greater but frequency is less. At higher speeds the reverse is true. The
greatest harm is caused by the amplitude of vibration which is the factor, most
destructive of instruments and which causes the most apprehension in the
patient and the greatest fatigue in the dentist.
By
increasing operating speeds, the amplitude and its effects are reduced and a
more satisfactory result is attained.
Vibration
waves are measured in cycles per second. It has been shown that rotation of
approximately 6000 rpm sets up a vibrational wave of approximately 100 cps. As
the rpm is increased the cps of the fundamental vibration wave are increased
until, at ranges of 100000 rpm, we have a vibration wave of 1600 cps. It has
been demonstrated that at wave of vibration of over 1300 cps, vibration is
practically imperceptible to the patient. The reason for this is not fully
understood, but there are two theories for this phenomenon.
(1) The
Wedensky inhibition phenomenon – frequency increased to a point where vibratory
perception diminishes due to failure to perceive vibration. This is because
Stimulation occurs during Refractory Period of Recovery.
(2) Vibratory
perception depends upon the product of the amount of stimulation (i.e. pressure)
multiplied by the frequency of stimulation necessary for a reaction. This is
called Chronaxie. As the speeds above 1,00,000 rpm, due to light pressure and
high speeds, chronaxie is attained, which is necessary for reaction.
Thus
it can be concluded that, the more the rpm, the less the amplitude, and the
greater the frequency. Vibratory perception will be lost in the ultra highspeed
range of 1,00,000 rpm or more.
Spread of pathogenic organisms by Ultra speed cutting procedures:
Atmospheric
contamination through the spread of oral organisms particularly from air
turbine action has been a concern of the dental profession for some time.
Dental procedures tend to expose the operator to infectious diseases. Recent
studies suggest that the extent of aerosol produced by air turbine may increase
the normal hazard. A report involving patients with pulmonary tuberculosis
cultures were demonstrated on all petri dishes exposed during cutting
procedures, with the heaviest concentration being at 2 feet in distances from
the patients mouth. This indicates that the dentist and his assistant are
exposed to a serious health hazard when operating with an ultra speed exposed
instrument on patients having such pathogens in their oral flora. When a
patient’s history suggests the existence of tuberculosis, pneumonia, influenza,
infections hepatitis or any infectious diseases including the common cold, a
protective face mask should be worn by both dentist and assistant. During all
ultrahigh speed cutting procedures, protective eye-glasses should be worm
routinely.
SUCK-BACK
PHENOMENON- The operation of the turbine
is switched off by closing the compressed-air valve abruptly. Then, owing to
its own kinetic energy, the turbine continues its rotation, so that the turbine
starts operating as an air pump. This causes a negative pressure in the area of
the turbine shaft. The negative pressure sucks air from the environment that
can be contaminated by aerosols of saliva and blood of the patient.
Size of cutting instrument and cutting speeds:
It
has been pointed out by Peyton, and Nelson that, the important factor of
increased operating speeds is the instrument surface speed in fact per minute
rather than the revolution per minute of the instrument.
The
larger the diameter of the cutting instrument, the slower the speed required at
the spindle. The specific phase in preparation of an abutment should determine
the size of the cutting instrument and the rpm that should be used. Employing
superspeed for all operations places unnecessary strain upon the patient and
equipment. If the same effect can be accomplished by using a larger instrument
at a lower speed, but still remaining above the threshold of perception, this
should be done. However, oversized cutting tools should not be used at super
speeds due to the difficulty of instrument control and accuracy of cutting.
VIBRATION
SYNDROME : the perception of vibration, pain, touch and
temperature deteriorates. The negative effect of local vibrations occurs within
the range 5-1400 hz, the most harmful being those below 16 hz. mechanical
vibrations arise because the various machines operating at the dentist’s
workplace contain moving parts. The main source is vibrating power-driven or
air-driven instruments, such as low- and high-speed handpieces as well as
ultrasonic instruments. The vibrations
emitted by these machines travel directly from the handles to the operator’s
hand. These are local vibrations.
Biologic response of dentin and pulp to high speed cutting:
Dentin: As the contents of the dentinal
tubules are in direct continuity with the odontoblasts, and pulp, cutting or
grinding the dentin causes a reaction in the pulp and this may lead to changes
in the dentin.
An early experimental investigation
into the effect of cavity preparation on the dentin and pulp was carried out by
Fish in 1932. He cut cavities in the teeth of dogs and left the cavities open
to the saliva. By sealing dyes into the pulp chambers of the treated teeth he
has shown that one of two reactions is produced in the dentin.
In
some cases there was sclerosis of the cut dentinal tubules which forms a
protective some sealing off the pulp from the injury and underneath this
region, there is a further growth of tubular dentine. These reactions are
produced by the stimulation of the odontoblasts. The other reaction that
resulted was the formation of dead tracts. With this lesion some or all of the
odontoblasts, that are in connection with the cut tubules die. On the pulpal aspect
of these tubules, hyaline mineralized barrier, secondary dentine is laid down,
thereby sealing the lesion from the pulp.
Pulp: The changes in the pulp have been
studied by Langeland and Morslard and Shovelton. They state that the damage to
the pulp is to a large extent due to the heat generated. They have shown that
when precautions are taken to minimize heat production by using burs rotated
slowly in a speed reducing handpieces, the only evidence of pulp damage was a
slight reduction of the odontoblast layer with the displacement of a small
number of odontoblasts into the dentinal tubules. When speeds upto 5,000 rpm
were employed, there was more extensive displacement of odontoblasts associated
with marked vacuolization of the odontoblast layer, and local hemorrhages may
be seen in the pulp. As the speed was increased, the changes became more
severe. When tooth reduction was done under a stream or spray of water, the
damage to the pulp was markedly reduced.
Pulp changes associated with tooth
reduction using the air abrasive technique have been studied by Kennedy and
using ultrasonic technique by Mitchell and Jenson. The changes in both the
cases are similar to those produced at the speeds of 5,000 rpm.
The
effects on the pulp of using high speed rotary instruments such as the air
turbine have been investigated by Marsland and Shovelton. The changes found are
no severe than those produced at lower operating speeds provided that adequate
cooling of the cutting instrument by water jet or air/water spray is ensured.
Alterations
in the hard tissues of tooth cut by air turbines have been observed. The enamel
over a wide area of crown may show minute cracks and the dentin shows altered
staining reactions as a result a local overheating.
RECENT
DEVELOPMENT:
‘ANTI-SUCK
BACK’- Planmeca compact dental
units, the turbine drive air is not shut off abruptly but controlled down by
allowing the driving air to decrease gradually. The software of the dental unit
will keep on supplying the drive air into the turbine according to carefully
chosen parameters. This way there is no possibility for the build-up of a
vacuum effect that would cause suck-back.
Ceramic bearings- no need of lubrication and more resistant to autoclave
sterlization.
Use of quartz rods instead of fibre-optic.
Easy-to-use push-button bur releases.
Swivel systems.
Titanium handpieces.
Smaller head size.
n ELECTRICAL
HIGH SPEED HANDPIECES
advantages are:
More power and torque than air turbine handpieces.
Better bur concentricity.
Less vibration and noise.
Broad, controllable speed ranges.
Forward/reverse option are available.
With appropriate attachments, one
system can be used for restorative, prosthodontics, prophies and endodontics.
Disadvantages are :
Heavier than air turbine.
More expensive.
Learning curve may be required.
Attachment heads not as small as the small-head air turbines.
May not be able to fully replace the air turbine.
Infection control concerns.
Discussion
It
is for more than 125 years, that rotary instruments have been in use, for tooth
reduction operations, in different forms, from a hand rotary instrument to
ultra sonic instruments, which have the rotational speeds ranging from very low
speeds in case of band rotary instruments to 5,00,000 rpm in case of air
bearing hand piece. These remarkable advances in the instruments have greatly
reduced fatigue in the operator because of the physical case of manipulation
and have considerably increased the comfort to the patient by reducing the
actual working time and pressures required for tooth reduction, thereby
minimizing the factors of heat generation and pain. Though high speed
techniques have been a born to the dental profession, they have their can
limitations. It is interesting to not that, in spite of considerable improvements
in tooth reduction procedures and the instruments used for the same, the
principles and biologic objectives have not changed.
These
improved methods of tooth tissue removal have a potential to damage the healthy
teeth and surrounding structures, if they are used without proper understanding
of their working and if they are used without taking proper care. Improper
handling of these modern equipment may also be different to the longevity and
working capacity of the instruments themselves.
For
successful and efficient use of those cutting tools, certain factors should be
given consideration. Heat that is generated, while the tooth tissue is being
removed must be kept, down to the minimum and at the sametime, whatever heat is
generated, must be eliminated as efficiently and as quickly as possible by
employing coolants, in any one of three forms commonly used i.e. water,
air/water spray or air alone. Simultaneously with the coolant, if water or
air/water spray is used, an efficient mechanism for remove of the water from
the oral cavity must be employed. Otherwise, the clinical procedure is delayed,
if the patient has to spit out the water, every now and then. By eliminating
the water evacuation equipment, we are losing one of the advantages of these
high speed instruments i.e. reduced working time for a particular preparation.
Use of efficient coolants, not only eliminate the heat generated, but at the
same time, keeps the operating area clean and free of any debris.
High
speed cutting methods have a further advantages in that, they reduce the
annoyance that may be caused to the patient, when low speeds are used with the
modern high speed cutting devices, the vibration produced is of a frequency
that is generally beyond the upper threshold of perception of the average
patient.
Pressures
that have to be employed in the use of high speeds are considerably reduced, in
comparison with those needed for low speeds.
Thus,
when the factors of pressure, temperature and vibration are kept within the
tolerance limits, the patient comfort is certainly improved.
Size
of the cutting tool to be used for particular tooth reduction procedure is an
important consideration, particularly while using high speeds. Oversized
cutting tools should be avoided, as they are difficult to control and at the
same time, the accuracy of tooth preparation on procedure is also adversely
affected.
Biologic
reactions of the tooth tissues, particularly dentin and pulp, should not be
over locked, when high speeds are employed for tooth reduction operation. These
responses have been studied by a number of people and they have shown that, the
response are not significantly different from those, when low speeds are used,
provided, effective coolants are employed.
Thus
it can be concluded that, high speed equipments for tooth reduction if used
with proper understanding and due care, provide definite advantages over the
conventional low speed cutting procedures. This fact places the high speed
devices at definitely a higher level as against their low speed counterparts.
Conclusions
1.
High speed cutting devices, if used
with a thorough understanding of their mechanism and due care to the biologic
integrity of teeth and surrounding structure, are a boon to dentistry.
2.
In the process of tooth reduction
using high speeds considerable amount of heat is generated and an effective
coolant is a must for preservation of tooth integrity and patient comfort.
3.
Vibration is increased with the
increase in speed, but it is beyond the threshold of prerception of the normal
human beings and hence not harmful.
4.
Biologic reactions of the dentin
and pulp, to high speed cutting, cannot be overlooked.
Summary
A
brief history of rotary instruments has been presented. A critical evaluation
of the high speed cutting devices, as to their advantages, disadvantages, and
precautions to be taken during their use, has been assessed at length. Biologic
reactions of dentin and pulp, to high speed cutting have been discussed in
brief.
Contents
I.
Introduction
II.
Review of Literature
a.
Advantages of high speeds
b.
Disadvantages of high speeds
c.
Types of high speed instruments
d.
Heat generation
e.
Coolants
f.
Vibration
g.
Spread of pathogenic organisms
h.
Size of cutting instruments and cutting speeds
i.
Biologic responses of dentin and pulp to high speed
cutting
III. Discussion
IV. Conclusion
V.
Summary
|
No comments:
Post a Comment